A plethysmographic sensor for monitoring volume changes in cardiovascular pathologies
Description
This paper presents a capacitive system capable
of performing leg volume change monitorization in patients
suffering from Heart Failure (HF) conditions. The body volume
evolution serves as a prognosis marker for this kind of patients,
such patients can benefit from a wearable monitorization system.
The proposal is based on a contactless capacitive wearable
structure implemented by a two-plate plane-parallel capacitor
geometry. The system exhibits sensitivity to leg volume change
and the sensor curves are provided. A 2.5x2.5cm capacitive
electrode design will generate capacitive values within the range
[1-2] pF. Acquisition of the capacitance value is performed via an
electronic differentiator implemented using op-amps, illustrating
good results in simulated volume analysis implemented using
pspice.
Abstract
Ministerio de Ciencia, Innovación y Universidades RTI2018-093512-B-C21Additional details