Published May 19, 2023 | Version v1
Publication

A method for the determination of counting efficiencies in gamma-spectrometric measurements with HPGe detectors

Description

In this paper a general method for y-ray efficiency calibration is presented. The method takes into account the differences of densities and counting geometry between the real sample and the calibration sample. It is based on the y-transmission method and gives the correction factorf as a function of Ey, the density and counting geometry. Altough developed for soil samples, its underlying working philosophy is useful for any sample whose geometry can be adequately reproduced.

Additional details

Identifiers

URL
https://idus.us.es/handle//11441/146464
URN
urn:oai:idus.us.es:11441/146464

Origin repository

Origin repository