Published April 22, 2025 | Version v1
Publication

Comparison of the performance of the Quantulus GCT 6220 and the Quantulus 1220 for the determination of 55Fe and 63Ni

Description

An experimental study of the response of the newer Quantulus GTC 6220 liquid scintillation spectrometer has been conducted with the objective of determining the performance of this counter for the determination of 55Fe and 63Ni in the environment. Its response has been compared to that of the well-established Quantulus 1220. For this, Guard Compensation Technology (GCT) was used to reduce the Lower Limit of Detection (LLD in what follows). Two scintillation cocktails, Hidex ProSafe HC+ and Revvity OptiPhase HiSafe 3 were compared in this work. The parameters employed for this assessment were background count rate, counting efficiency, and LLD.

Additional details

Identifiers

URL
https://hdl.handle.net/11441/171909
URN
urn:oai:idus.us.es:11441/171909

Origin repository

Origin repository
USE