Published October 9, 2018
| Version v1
Publication
A first order incremental analog to digital converter based on continuous time circuits
Description
In this paper, an incremental Analog-to- Digital Converter (ADC) designed as part of the signalconditioning circuitry for tissue impedance measurement system is presented. Continuous-time design techniques has been used and a modified implementation of the conversion algorithm, with respect to its discrete-time counterpart, has been developed. To reduce the influence of the no-idealities, analog and digital corrections have been also implemented. A prototype in 0.8μm CMOS technology has been fabricated and tested. Experimental results are reported.
Additional details
- URL
- https://idus.us.es/handle//11441/79236
- URN
- urn:oai:idus.us.es:11441/79236
- Origin repository
- USE