Published 2014
| Version v1
Publication
Large area TES spiderweb bolometer for multi-mode cavity microwave detect
Description
Large area spiderweb bolometers of 8 mm diameter and a mesh size of 250 Î1/4m are fabricated in order to couple with approximately the first 20 modes of a multimode EM cavity at about 140 GHz. The sensor is a Ti/Au/Ti 3 layer TES with Tc tuned in the 330-380 mK and 2 mK transition width. We describe the detector design and the fabrication process, early TES electro-thermal measurements. We also report optical coupling measurement and show the multimode coupling.
Additional details
- URL
- http://hdl.handle.net/11567/897565
- URN
- urn:oai:iris.unige.it:11567/897565
- Origin repository
- UNIGE