Evaluation of three automated systems for susceptibility testing of enterobacteria containing qnrB, qnrS, and/or aac(6′)-Ib-cr
Description
The accuracy of the MicroScan WalkAway, BD Phoenix, and Vitek-2 systems for susceptibility testing of quinolones and aminoglycosides against 68 enterobacteria containing qnrB, qnrS, and/or aac(6′)-Ib-cr was evaluated using reference microdilution. Overall, one very major error (0.09%), 6 major errors (0.52%), and 45 minor errors (3.89%) were noted.
Abstract
Ministerio de Sanidad y Consumo y FEDER PI050690
Abstract
Spanish Network for Research in Infectious Diseases REIPI RD06/0008
Abstract
Junta de Andalucía P07-CTS-02908 PI-0282-2010
Additional details
- URL
- https://idus.us.es/handle/11441/63357
- URN
- urn:oai:idus.us.es:11441/63357
- Origin repository
- USE